Continuation of Guide Evaluations:
-> All runs this series at beam height.
-> Timing for Ge gamma/CN measurements: 104s GVopen, beam ON | 51s GVclosed, beam ON
-> Ge detector always pushed up until it touches test guide.
-> Transmission measurements with an upstream pinhole as before, TPX on downsatream side,
Ni on upstream side.
-> The bottling measurements in this series use the same upstream aperture as used previously,
but use a 21/64" or 15/32" dia hole in a Ni foil as the downstream aperture (~twice or ~four times
the area of the previous aperture). Timing is as before, 120s GVopen, BEAMon and 120s GVcolsed, BEAMoff.
Viewed looking downstream, the upstream aperture is at 9 o'clock and the downstream aperture
is at 3 o'clock.
sdaq fdaq Main Mon
run# run# Guide Measurement Type Counts Counts Comments
------------------------------------------------------------------------------------------------------
435 11618 2.75OD EP SS Ge gamma/CN -- -- logger died ~30min into run
435 11620 2.75OD EP SS Ge gamma/Cn -- -- restarted fdaq for remainder of run
438 11621 2.75OD EP Cu 39.53" transmission 43629 41830 consistency check of earlier run (Cu guide 0.4" past Al adapter, UCN flow in same dir through tube), trans ~92.5%
439 11622 2.75OD DLC Cu 39.37" transmission 48000 48440 consistency check of earlier run (Cu guide 0.4" past Al adapter, UCN flow same dir through tube), trans ~87.8%
441 11623 2.75OD DLC Cu 39.37" bottling (21/64") -- -- cycles: 16 ld cnts: 19549 eq rate: 249.4 dr cnts: 6614 ld tc: 44.7 dr tc: 27.4 mon cnts: 9558
443 11625 2.75OD DLC Cu 39.37" bottling (15/32") -- -- cycles: 16 ld cnts: 10239 eq rate: 106.0 dr cnts: 1411 ld tc: 23.6 dr tc: 13.0 mon cnts: 5626 -> number for last 10 cycles only
444 11626 3" H-DLC SS after D2 plasma Ge gamma/CN -- --
|